חיפוש
סגור את תיבת החיפוש

FRANKONIA – CHC – Compact Hybrid Anechoic Chamber

Frankonia CHC – Pre-compliant 3m Compact EMC Chamber

Overview Frankonia CHC – Compact Hybrid Chamber
EMI Pre-compliant
EMS IEC/EN 61000-4-3, full compliant
Test distance 3,0 m
Volume ø1,2 m
Special Compact immunity test chamber

The CHC is Frankonia’s compact hybrid chamber solution at 3,0 m measuring distance with a Quiet Zone (QZ) of ø1,2 m. The CHC is an upgradable solution and can transform from a semi anechoic chamber configuration with ground plane to a fully anechoic chamber configuration with floor absorbers. It is an optimal solution for both pre-compliance emission tests and full compliance immunity tests at 3,0 m measuring distance.

The CHC can be delivered optionally as a semi-anechoic chamber (with ground plane) or as a fully anechoic chamber (floor absorbers on the whole floor surface).

For immunity tests, floor absorbers are required in each case between transmitting antenna and EUT and are therefore included in the standard delivery volume.

Typical Specifications

Chamber

CHC

CHC-L 

External dimension

7.355 x 3.755 x 3.300 mm (LxWxH)

8.255 x 3.755 x 3.300 mm (LxWxH)

Quiet Zone

ø1,2 m

Measuring distances

3,0 m

Load capacity

up to 2,0 tons; special turntable systems possible

Frequency range

9 kHz/ 30 MHz to 18 GHz (option 40 GHz)

Typical Product Standards

Emission

 Pre-compliance from 30 MHz to 1 GHz

Immunity

 IEC/EN 61000-4-3

Typical Verification Standards

Emission

 CISPR 16-1-4

Immunity

 IEC/EN 61000-4-3

Performance:

SHIELDING EFFECTIVENESS (SE)
Acc. to EN 50147-1 from 10kHz to 18GHz the following values are guaranteed: see table paragraph 1

EMISSION – pre-compliance
According to CISPR 16-1-4
Test distance: 3,0m
Frequency range: 30MHz- 200MHz (pre-compliance)
Frequency range: 200MHz- 1GHz (compliance)
Quiet Zone: Ø1,2m (2m H)

 

IMMUNITY – pre-compliance
According to IEC/EN 61000-4-3/ ISO 11452-2
Test distance: 3,0m
Frequency range: 80MHz- 18GHz
Uniform area: 0,5m x 0,5m
Guaranteed FU: -0/+6dB for 12 of 16 test points (75%)

 

Outside shielding dimensions (LxWxH): 7,355 x 3,755 x 3,300m

Shielding and structure

Quantity

Unit

Shielding modules PAN type

1

Set

Ceiling/ wall reinforcement

1

Set

Rubber membrane under the chamber for humidity protection

1

Set

Absorber ferrite tiles F006
Floor lined with F006 (7,2m²) for EMS (FU) 1 Set
Walls and ceiling lined with F006 1 Set
Rail substructure for ferrite tile plates 1 Set
Absorber Frankosorb® pyramids – fire class B2 DIN 4102/ DIN EN 13501-1 (2010-01)
Floor with H600 B2 (7,2m²) for EMS (FU)

1

Set

Small wall behind the turntable lined with H450 B2

1

Set

Side walls partially lined with H450 B2 (approx. half of the chamber)

1

Set

Ceiling partially lined with H450 B2 (approx. half of the chamber)

1

Set

 

Personal door type 2011 – External leaf painted RAL 9002 (light grey)

SLD 10/20/M Personal door with clear opening 1013 x 2043mm with manual latching, prepared for receiving absorbers

1

Pc

Door stopper

1

Pc

Manual wooden ramp, load capacity 200kg

1

Pc

Penetration panels, feed throughs & connectors
Penetration panel (PP) 400 x 400mm, including C-frame

1

Pc

N connector

4

Pcs

N+BNC adaptor

4

Pcs

FSMA connector

4

Pcs

Floor connection panels & connectors
Connection panel (CP) 600 x 600mm – automatic with jacks

2

Pcs

N connector

4

Pcs

BNC connector

4

Pcs

FSMA connector

4

Pcs

 

 

Power filter and grounding point
Insertion loss high performance filters: 100dB from 14kHz to 40GHz
Power filter 2x6A – 250V AC – 50/60Hz for the chamber

1

Pc

Power filter 2x32A – 250V AC – 50/60Hz for the EUT

1

Pc

Electrical installation
Electrical installation

1

Set

Lighting Halogen 230W

4

Pcs

Shielded access box to the electrical distribution box with access from inside

1

Pc

Electrical ducts in raised floor

1

Set

Socket: 1phase 60Hz 32A (US type)

1

Pc

Socket: Duplex single phase 16A (US type)

3

Pcs

Emergency lighting low noise with 3 lights and outside UPS with filter

1

Set

Honeycombs
Honeycombs 400 x 400mm – 18GHz

2

Pcs

Floor
Height: 150mm
Raised floor modular standard up to 500 kg/m² with hot galvanized ground plane 2,0mm 1 Set

 

Standard Equipment
· 1 access door, 1.013 x 2.043 mm
· 1 honeycomb insert for ventilation
· 1 mains filter 250 VAC, 2 x 16 A
· 1 penetration panel
· Feed-throughs: 4 x “N”, 4 x “BNC” and 1 x for fibre optics
· Electric installation
· Illumination
· Raised floor
· Absorber lining

Options
· 3 phase mains filter
· Signal and / or data line filter
· Fan · Antenna tripod
· Video and / or audio system
· Verification of the chamber
· Measuring equipment
· Other extras

fOR MORE DETAIL, PLEASE DOWNLOAD THE FOLLOWING PDF FILE

Frankonia Absorber Overview Data Sheet Frankonia CHC – Compact Hybrid Chamber, 3m test distance Data Sheet

 

Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Interfaces for Franikonia Anechoic Chambers Data Sheet