חיפוש
סגור את תיבת החיפוש

FRANKONIA – FAC-3L Fully Anechoic 3M EMC Chamber

Frankonia FAC-3 L Fully Anechoic 3m EMC Chamber for floor-standing EUT’s

Overview FAC-3 L
EMI CISPR 16-1-4 and IEC/EN 61000-4-22, full compliant
EMS IEC/EN 61000-4-3 and IEC/EN 61000-4-22, full compliant
Test distance 3m
Volume ø1,5 m (H=2,0 m)
Special Free-space test environment for table-top and floor-standing EUT’s

The FAC-3 L is the extended version of Frankonia’s fully anechoic chamber solution at 3,0m measuring distance for EMC tests on table top positioned EUT’s as well as on floor-standing EUT’s. It is designed as full compliant chamber for measurements under free-space conditions and is based on CISPR 16-1-4 as a test site without ground plane. Without the reflections from the floor, a height scan is no longer necessary. With its specific requirements for the test site, Frankonia’s FAC-3 is supremely prepared to meet our customers’ demands.

Features

• Test site for table top and floor-standing EUT’s

• Full compliant EMI/EMS

• Cost effective solution for free-space measurements

• Compact chamber design with advanced absorber lining with long-lasting Frankosorb® absorbers (Frankonia technology)

• Turnkey solution

Typical Product Standards

Emission

• CISPR 14

• CISPR 32

• ETSI

Immunity

• IEC/EN 61000-4-3

Typical Verification Standards

Emission

• CISPR 16-1-4

• IEC/EN 61000-4-22

Immunity

• IEC/EN 61000-4-3

• IEC/EN 61000-4-22

TYPICAL SPECIFICATIONS

Chamber FAC-3 L
External dimensions 9.380 x 5.780 x 5.550 mm (LxWxH)
Quiet Zone ø1,5 m (H=2,0 m); floor standing EUT tests
Turntable ø1,5 m
Measuring distances 3,0 m
Load capacity up to 2,0 tons; special turntable systems possible
Frequency range 9 kHz/ 30 MHz to 18 GHz (option 40 GHz)

Performance & Compliance

• Full compliant emission (EMI) according to CISPR 16-1-4

Deviation FS NSA ±3,5 dB (30 MHz to 1 GHz) – guaranteed

Deviation FS SVSWR +5,5 dB (1 GHz to 18 GHz) – guaranteed

• Full compliant immunity (EMS) according to IEC/EN 61000-4-3

Deviation FU 0 dB/+6 dB at 75 % of 16 measuring points (26/ 80 MHz to 18 GHz) – guaranteed

• Full compliant immunity (EMS) and emission (EMI) according to IEC/EN 61000-4-22 Standard

Deviation SdB c ≤ 1,8 dB

Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Interfaces for Franikonia Anechoic Chambers Data Sheet