Frankonia SAC-10/P Pyramid Semi Anechoic 10m EMC Chamber with Long-pyramid P2400 Absorbers
Overview | SAC-10/P Pyramid |
EMI | CISPR 16-1-4 and ANSI C63.4, full compliant |
EMS | IEC/EN 61000-4-3, full compliant |
Test distance | up to 10,0m |
Volume | ø3,0 m up to ø6,0 m |
Special | P2400 Long-pyramid Absorbers |
The SAC-10/P chambers are Frankonia’s full compliant and customizable EMC testing solutions at 10,0m measuring distance offering various sizes of the Quiet Zone (QZ). Due to the high grade of customization reflecting the demands of our customers, this semi-anechoic chamber is adaptable in size and offers several configuration possibilities. The innovative concept with its impressionable absorber layout achieves exceptional performance for emission and immunity testing.
Features
• Unique Frankosorb® P2400 Pyramid Absorbers
• Full compliant EMI/EMS
• Highly customizable solution for any kind of EMC testing and limitless integration of individual applications
• Adjustable chamber size, characteristics and configuration due to different EUT requirements
• Advanced absorber lining with long-lasting Frankosorb® absorbers (Frankonia technology)
• Specialized for ‘out-of-the-range’ EMC test environments
• Turnkey solution
Typical Product Standards
Emission
• CISPR 11
• CISPR 12
• CISPR 14
• CISPR 15
• CISPR 25
• CISPR 32
• MIL-STD 461
Immunity
• IEC/EN 61000-4-3
• ISO 11451
• ISO 11452
• MIL-STD 461
• RS 101, RS102, RS103
Typical Verification Standards
Emission
• CISPR 16-1-4 and/or ANSI C63.4
Immunity
• IEC/EN 61000-4-3
Chamber Type |
SAC-10-3/P |
SAC-10-4/P |
SAC-10-5/P |
SAC-10-6/P |
External dimension |
21.680 x 13.730 x 8.550 mm (LxWxH) |
21.680 x 13.730 x 8.550 mm (LxWxH) |
23.480 x 16.580 x 9.000 mm (LxWxH) |
24.980 x 17.180 x 9.000 mm (LxWxH) |
Quiet Zone at 3,0m |
ø3,0 m |
ø3,0 m |
ø3,0 m |
ø3,0 m |
Quiet Zone at 5,0m |
ø3,0 m |
ø4,0 m |
ø5,0 m |
ø5,0 m |
Quiet Zone at 10,0m |
ø3,0 m |
ø4,0 m |
ø5,0 m |
ø6,0 m |
Load capacity |
up to 80,0 tons; special turntable systems and dynamometer integration possible |
|||
Frequency range |
9 kHz/ 30 MHz to 18 GHz (option 40 GHz) |
Performance & Compliance
• Full compliant emission (EMI) according to CISPR 16-1-4 and ANSI C63.4
Deviation NSA ±3,5 dB (30 MHz to 1 GHz) – guaranteed
• Full compliant immunity (EMS) according to IEC/EN 61000-4-3
Deviation FU 0 dB/+6 dB at 75 % of 16 measuring points (26/ 80 MHz to 18 GHz) – guaranteed