חיפוש
סגור את תיבת החיפוש

FRANKONIA – SAC-5 SEMI ANECHOIC EMC CHAMBER

Frankonia SAC-5 Square Semi Anechoic 5m EMC Chamber with traditional square design

Overview SAC-5 Square
EMI CISPR 16-1-4 and ANSI C63.4, full compliant
EMS IEC/EN 61000-4-3, full compliant
Test distance 5,0 m & 3,0 m
Volume up tp ø3,0 m
Special Traditional square design

The SAC-5 Square is Frankonia’s full compliant EMC testing solution at 5,0 m measuring distance with a Quiet Zone (QZ) up to ø3,0 m. It is adapted for full compliant emission and immunity testing with a traditional square design.

 

Features

• Cost-efficient and high-performance solution for a 5,0 m measuring distance

• Full compliant EMI/EMS

• Adapted steel structure and optimized RF-shielding with traditional rectangular design

• Upgradeable for E-Drive (load machine, BlueBox, battery test system)

• Ingenious absorber lining with Frankosorb®

• Outstanding performance with long-lasting Frankosorb® non-combustible absorbers

• Turnkey solution

Typical Product Standards

Emission

• CISPR 11

• CISPR 14

• CISPR 15

• CISPR 25

• CISPR 32

• MIL-STD 461

Immunity

• IEC/EN 61000-4-3

• ISO 11452

• MIL-STD 461

Typical Verification Standards

Emission

• CISPR 16-1-4 and/or ANSI C63.4

Immunity

• IEC/EN 61000-4-3

Typical Specifications

Chamber SAC-5 Square
External dimensions 12.680 x 8.180 x 6.000 mm (LxWxH)
Quiet Zone ø3,0 m
Measuring distances 5,0 m and 3,0 m
Load capacity up to 5,0 tons; special turntable systems possible
Frequency range 9 kHz/ 26 MHz to 18 GHz (option 40 GHz)

Performance & Compliance

• Full compliant emission (EMI) according to CISPR 16-1-4 and ANSI C63.4

Deviation NSA ±3,5 dB (30 MHz to 1 GHz) – guaranteed

Deviation SVSWR +5,5 dB (1 GHz to 18 GHz) – guaranteed

• Full compliant immunity (EMS) according to IEC/EN 61000-4-3

Deviation FU 0 dB/+6 dB at 75 % of 16 measuring points (26/ 80 MHz to 18 GHz) – guaranteed